HPLC Troubleshooting: Point-by-Point Guide
• Check injection volume – Was the sample injected correctly?• Verify sample preparation – Ensure correct solvent and concentration.• Inspect flow rate – Ensure mobile phase is flowing at correct rate.• Confirm detector settings – Wavelength, range, and mode should match the compound.
2.
• Sample might be diluted.• Possible sample degradation before injection.• Incorrect injection volume (too small).
3.
• Sample over-concentrated.• Injection volume too high.• Integration error in software (baseline not properly set).
4.
• Caused by air bubbles in the system.• Detector lamp aging (especially UV lamps).• Contaminated mobile phase or solvents.
5.
• Temperature fluctuations during run.• Mismatch between sample and mobile phase.• Detector instability or warm-up issues.
6.
• Mobile phase contamination (use fresh solvents).• Sample carryover from previous runs.• Dirty injector or loop (perform cleaning).
7.
• Incomplete sample dissolution.• Column overload or partial blockage.• Inappropriate solvent compatibility.
8.
• Aged or damaged column.• Poor sample solubility.• Flow rate too slow.
9.
• Problem with column integrity or packing.• Sample solvent mismatch with mobile phase.• Residual silanol activity (use end-capped columns or buffer).
10.
• Column overloading.• Too high injection volume.• Highly concentrated sample.
11.
• Mobile phase composition changed.• Flow rate adjustments during run.• Temperature variations.
12.
• Mobile phase not mixed consistently.• Pump inconsistencies or mechanical issues.
13.
• Blocked column, frit, or inline filter.• Viscous mobile phase or buffer precipitation.• Tubing clogged or narrowed.
14.
• Leak in system (fittings, pump, or injector).• Column not properly installed or frit missing.
15.
• Check pump head, injector, and tubing connections.• Tighten any loose fittings.
16.
• Degas mobile phase (ultrasonic or vacuum).• Prime the pump before use.
17.
• Air in flow cell – degas mobile phase.• Dirty or old UV lamp – check lamp age/hours.• Unstable electrical power supply.
18.
• Inconsistent injection volumes.• Errors in sample preparation or handling.• Syringe or injector malfunction.
19.
• Presence of isomers or closely related compounds.• Column not efficient – check particle size and type.
20.
• Air bubbles in pump head.• Leaks in suction line or fittings.• Mobile phase degassing insufficient.